Axiom Artis DFC (2014-02-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Serial number: 35051
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
Axiom Artis DFC (2014-02-06)7412807

Introduction

Recalled Products

Axiom Artis DFC 

Reason

During a technical review, it was found that Axiom Artis systems with a certain serial number are using a flat detector cooling unit with inappropriate connectors that may possibly leak. This might cause a system failure which may possibly result in a drop out of the gantry system. This update does only affect systems of a specific production series.