Cognis Cardiac Defibrillators (N106 and N107) and Teligen Implantable Defibrillators (N102 and E110) (2013-08-29) Aug 29, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Brand name Companies Model or catalog number Not applicable D. Teligen Implantable Defibrillator E110 Manufacturer Cardiac Pacemakers Incorporated 4100 Hamline Avenue North St. Paul 55112 UNITED STATES E110 Not applicable B. Cognis Cardiac Defibrillator N107 Manufacturer Cardiac Pacemakers Incorporated 4100 Hamline Avenue North St. Paul 55112 UNITED STATES N107 Not applicable C. Teligen Implantable Defibrillator N102 Manufacturer Cardiac Pacemakers Incorporated 4100 Hamline Avenue North St. Paul 55112 UNITED STATES E102 Not applicable A. Cognis Cardiac Defibrillator N106 Manufacturer Cardiac Pacemakers Incorporated 4100 Hamline Avenue North St. Paul 55112 UNITED STATES N106
Recalled Products
Cognis Cardiac Defibrillators (N106 and N107) and Teligen Implantable Defibrillators (N102 and E110)
Performance of a low voltage capacitor in the subset of devices may be compromised over time, causing increased current drain that can lead to premature battery depletion.