Plegiox cardioplegia heat exchanger recall (2014-01-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
BO-H 65960
Manufacturer
Maquet Cardiopulmonary AG
Kehler Str. 31
Rastatt
76437
GERMANY
F. Cardioplegia Set with Softline Coating92118350
BO-HQV 46901
Manufacturer
Maquet Cardiopulmonary AG
Kehler Str. 31
Rastatt
76437
GERMANY
D. Custom Pack with Softline92115322
H 55807
Manufacturer
Maquet Cardiopulmonary AG
Kehler Str. 31
Rastatt
76437
GERMANY
E. Heart Lung Pack with Rotaflow and Plegiox92112815
BEQ-HQV 46900
Manufacturer
Maquet Cardiopulmonary AG
Kehler Str. 31
Rastatt
76437
GERMANY
B. Standard Pump Pack Bioline92109788
BO-HQV 48707
Manufacturer
Maquet Cardiopulmonary AG
Kehler Str. 31
Rastatt
76437
GERMANY
C. Heart Lung Perfusion Pack with Softline92120338
BE-HQV 48707
Manufacturer
Maquet Cardiopulmonary AG
Kehler Str. 31
Rastatt
76437
GERMANY
A. Perfusion Pack Bioline92111319

Introduction

Recalled Products

A. Perfusion Pack Bioline
B. Standard Pump Pack Bioline
C. Heart Lung Perfusion Pack with Softline
D. Custom Pack with Softline
E. Heart Lung Pack with Rotaflow and Plegiox
F. Cardioplegia Set with Softline Coating

Reason

The desired cardioplegia temperature may not be obtained when the Plegiox cardioplegia heat exchanger (Plegiox) is used in conjunction with a heater/cooler unit. Isolated field reports have indicated that a user could not reach a desired low cooling temperature of Cardloplegic solution, as the measured value of the cardioplegia temperature by the probe on the Plegiox did not match the set temperature on the heater/cooler unit. This variation has been attributed to the position of the temperature probe at the port on the Plegiox, which may lead to temperature readings differing to a varying degree from the real temperature of the Cardioplegic solution.  The discrepancy between cardioplegia temperature measured at the probe of the Plegiox and set water temperature on the heater/cooler unit may appear greater than typical during clinical use of the heater/cooler system.