I-stat 1 Analyzer system - G3+ Cartridge (2014-01-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
03P78-25I-stat 1 Analyzer system - G3+ Cartridge 
Manufacturer
Abbott Point of Care Inc.
100 and 200 Abbott Park Rd.
Abbott Park
60064
Illinois
UNITED STATES
N13183

Introduction

Recalled Products

I-stat 1 Analyzer system - G3+ Cartridge 

Reason

Abbott point of care Inc. (APOC) has determined that I-stat G3+ cartridges from cartridge lot number N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results. Abbott point of care has determined that some individual patient results generated with this cartridge lot have the potential to exhibit incorrectly elevated pco2 and depressed pH results. Internal studies have demonstrated that PCO2 results can be elevated by 10 to 25 mm HG, pH results can be depressed by 0.10 to 0.15. All other G3+ cartridge lots are unaffected by this issue.