Sigma Spectrum Infusion System with MDL (2014-02-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
35700
Manufacturer
Baxter Healthcare Corp.
711 Park Ave.
Medina
14103
New York
UNITED STATES
Sigma Spectrum Infusion System with MDLMore than 10 numbers, contact manufacturer 

Introduction

Recalled Products

Sigma Spectrum Infusion System with MDL

Reason

Baxter has received complaints for repeated system error 322 "link switch error (low)" occurrences with the Sigma Spectrum Infusion System. This error may occur when there is a miscommunication with the two internal switches that detect door closure. If system error 322 occurs during infusion, the Sigma Spectrum Infusion System stops infusing. An audible alarm and visual alarm will also occur.