Architect I system - Intact PTH Reagent Assay (2014-02-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
8K25-01
Manufacturer
Abbott GmbH and Co. K.G.
Max Planck Ring 2
Wiesbaden
65205
GERMANY
Architect I system - Intact PTH Reagent AssayAll lots

Introduction

Recalled Products

Architect I system - Intact PTH Reagent Assay

Reason

A performance shift was observed in the architect intact PTH assay that has the potential to generate falsely elevated results on patient samples. Results generated in a study using current in-date reagent and calibrator lots, demonstrated a magnitude of shift relative averaging approximately 13 to 45 percent when compared to package insert data. The shift was observed across the full analytical range of the assay.