Kodak Directview DR 7500 Dual Detector System (using version 5 software)
75000150
Introduction
Recalled Products
Kodak Directview DR 7500 Dual Detector System (using version 5 software)
Reason
Carestream has identified that if the system user interrupts the daily switch off-set refresh detector calibration by switching the active detector from one Bucky to another (e.g. table to the wall), and cancels detector calibration within 30 seconds of initiating this process, the previous patient image will be displayed on the console with the current patient's name. The likelihood of this issue occurring is improbable, and there have not been any reported injuries or misdiagnosis as a result.