Health Products

Heartstart MRX Defibrillator / Monitor (2013-04-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
M3535A
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
Heartstart MRX Defibrillator/MonitorAll lots

Introduction

Recalled Products

Heartstart MRX Defibrillator/Monitor

Reason

Philips Healthcare has become aware that a Philips Heartstart MRX monitor/defibrillator operating on battery power only that is used in-hospital or in pre-hospital environments may shutdown without warning if exposed to elevated levels of electromagnetic interference from other radio frequency (RF) energy sources and continuous repetitive power line bursts. Examples of other equipment that can cause interference are other medical devices, cellular products, information technology equipment and radio/television transmissions. Philips Healthcare will be providing all affected customers with an IFU addendum titled, Heartstart MRX electromagnetic compatibility.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime