Health Products

Heartstart MRX Defibrillator/Monitor (2014-02-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
M3535A
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
Heartstart MRX Defibrillator/MonitorAll lots

Introduction

Recalled Products

Heartstart MRX Defibrillator/Monitor

Reason

Philips has become aware that a component of the MRX processor board may be susceptible to damage from electrostatic discharge (ESD). Damage to these components can disrupt ECG and SPO2 functionality leading to an inability to perform:
  • demand mode pacing
  • 12 lead ECG acquisition and analysis
  • ECG monitoring via leads
  • pulse oximetry (SPO2)


The presence of this condition will be displayed on the ready for use (RFU) indicator as a solid red x with periodic audio chirp.  Defibrillation, fixed mode pacing, ECG monitoring via pads/paddles, and other monitoring functions are not impacted by this issue.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime