Philips has become aware that a component of the MRX processor board may be susceptible to damage from electrostatic discharge (ESD). Damage to these components can disrupt ECG and SPO2 functionality leading to an inability to perform:
demand mode pacing
12 lead ECG acquisition and analysis
ECG monitoring via leads
pulse oximetry (SPO2)
The presence of this condition will be displayed on the ready for use (RFU) indicator as a solid red x with periodic audio chirp. Defibrillation, fixed mode pacing, ECG monitoring via pads/paddles, and other monitoring functions are not impacted by this issue.