Health Products

XVI and XVI R 4.5 (X-Ray Volume Imaging Systems), and XVI R 4.5 System - Upgrade Kit (2013-11-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
MRT 14921
Manufacturer
Elekta Limited
Linac House, Fleming Way
Crawley, Sussex
RH10 9RR
UNITED KINGDOM
B. XVI R 4.5 (X-Ray Volume Imaging System)Version R 4.5
MRT 13221
Manufacturer
Elekta Limited
Linac House, Fleming Way
Crawley, Sussex
RH10 9RR
UNITED KINGDOM
C. XVI R 4.5 System - Upgrade KitVersion R 4.5
MRT 9271
Manufacturer
Elekta Limited
Linac House, Fleming Way
Crawley, Sussex
RH10 9RR
UNITED KINGDOM
A. XVI (X-Ray Volume Imaging System)Version R 4.2.1

Introduction

Recalled Products

A. XVI (X-Ray Volume Imaging System)
B. XVI R 4.5 (X-Ray Volume Imaging System)
C. XVI R 4.5 System - Upgrade Kit

Reason

Erroneous ASU table move in height direction may cause a collision when an external beam limiting device (BLD) (for example - Apex) is used.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime