A. Proteus XR/A General Purpose Radiographic System
Manufacturer
GE Hualun Medical Systems Co. Ltd.
No.1 Yong Chang North Road,, Beijing Economic Technological, Development Zone,
Beijing
100176
CHINA
00000094745HL7
Introduction
Recalled Products
A. Proteus XR/A General Purpose Radiographic System
B. Precision 500D System
Reason
Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams are improperly displayed. If the user continues to attempt to select one of the exams when this issue has occurred the system will update the screen, and in doing so potentially select a different exam/patient. If the user does not notice this change and continues to perform the exam, the images will be placed in an incorrect patient file.