Plum Lifecare 5000, Plum XL, and Plum XL 3 Infusion Systems and Plum XLM with Data Port (2014-01-27) Jan 27, 2014 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type I
Lot or serial number Companies Brand name Model or catalog number 11859 Manufacturer Hospira Inc. 275 North Field Drive Lake Forest 60045 Illinois UNITED STATES F. Plum XLM with Data Port All lots 11781 Manufacturer Hospira Inc. 275 North Field Drive Lake Forest 60045 Illinois UNITED STATES D. Plum XL 3 Multi-Line Infusion System All lots 11845 Manufacturer Hospira Inc. 275 North Field Drive Lake Forest 60045 Illinois UNITED STATES E. Plum XL 3 Micro/Macro Infusion System All lots 11555 Manufacturer Hospira Inc. 275 North Field Drive Lake Forest 60045 Illinois UNITED STATES B. Plum XL All lots 11846 Manufacturer Hospira Inc. 275 North Field Drive Lake Forest 60045 Illinois UNITED STATES C. Plum XL Micro/Macro Infusion System All lots 2507 Manufacturer Hospira Inc. 275 North Field Drive Lake Forest 60045 Illinois UNITED STATES A. Plum Lifecare 5000 Infusion System All lots
Recalled Products
Plum Lifecare 5000 Infusion System
Plum XL
Plum XL Micro/Macro Infusion System
Plum XL 3 Multi-Line Infusion System
Plum XL 3 Micro/Macro Infusion System
Plum XLM with Data Port
The door roller assembly on the Plum Lifecare 5000 series and Plum XL Infuser families has the potential to break which can lead to possible unrestricted flow and/or overdelivery when the door of the pump is opened.