Health Products

Gemstar Pumps (2014-02-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
13000-XX-YY
Manufacturer
Hospira Inc.
275 North Field Drive
Lake Forest
60045
Illinois
UNITED STATES
Gemstar PumpsAll serial numbers 

Introduction

Recalled Products

Gemstar Pumps

Reason

In Gemstar (Phase 3) infusion pumps, list numbers 13000 and 13150, the connection between the beeper subassembly and the pump may fail. The Gemstar infusion pump will identify this failure during the "self-test" while powering up which will result in a beeper error (X100). This beeper error (X100) is a service alarm that places the pump in an inoperable mode and requires service before it can be returned to service.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime