Advia Centaur XP System - Class 2, 3, and 4 (2014-02-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
078-A010-07
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
B. Advia Centaur XP System - Class 3More than 10 numbers, contact manufacturer
078-A010-07
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
C. Advia Centaur XP System - Class 4More than 10 numbers, contact manufacturer
078-A010-07
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
A. Advia Centaur XP System - Class 2More than 10 numbers, contact manufacturer

Introduction

Recalled Products

A. Advia Centaur XP System - Class 2
B. Advia Centaur XP System - Class 3
C. Advia Centaur XP System - Class 4

Reason

When the Advia Centaur XP Immunoassay system instrument cover is raised, the cover is supported by a gas spring attached at the middle of the cover. Over time, the gas spring may lose its effectiveness and fail to support the cover in its full or partially open position. This may lead to the cover falling during maintenance procedures.