Health Products

Heartstart FR3 Defibrillator (2013-02-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
861388
Manufacturer
Philips Medical Systems
22100 Bothell Everett Hwy.
Bothell
98021
Washington
UNITED STATES
Heartstart FR3 DefibrillatorAll lots

Introduction

Recalled Products

Heartstart FR3 Defibrillator

Reason

Philips has become aware that in a limited number FR3 Automated External Defibrillators (AED), the printed circuit board (PCB), which contains the circuitry that supplies energy to the device, could become separated from the electrical contact points that hold the board in place. In the event the PCB becomes separated from the electrical contact points, the AED may not be ready to deliver a shock in the event of an emergency.