Health Products

Phillips Heartstart HS1 Onsite Defibrillator (2014-02-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
M5066A-ABA
Manufacturer
Philips Medical Systems
22100 Bothell Everett Hwy.
Bothell
98021
Washington
UNITED STATES
Phillips Heartstart HS1 Onsite DefibrillatorAll lots

Introduction

Recalled Products

Phillips Heartstart HS1 Onsite Defibrillator

Reason

Philips has become aware that in a limited number of Heartstart Home and Onsite (HS1) Automated External Defibrillators (AED), one or both contact pins may be contaminated by residue from the soldering process. This contamination could prevent an adequate connection between the pads cartridge and the aed. If poor contact between the pads cartridge and aed occurs, the device may be unable to deliver therapy in an emergency.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime