Heartstart MRX Defibrillator / Monitor (2014-02-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
M3535AHeartstart MRX Defibrillator / Monitor
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
All lots

Introduction

Recalled Products

Heartstart MRX Defibrillator / Monitor

Reason

Philips has become aware that when used in hospital transport and pre-hospital (EMS) environments, the connection between the pads/paddles therapy cable (including pads CPR therapy cable, external paddles cable, and internal paddles cable/adapter) and the MRX therapy connection port could experience high levels of stress causing accelerated wear. Wear of the connection could prevent the device from sensing that the pads/paddles therapy cable is connected. This wear also could cause the MRX to inappropriately identify the pads therapy cable, external paddles, or internal paddles.  Pads/paddles therapy cable and therapy connection port wear could pose a risk for patients and/or caregivers.