Health Products

Elecsys AFP and Elecsys Total PSA (2014-02-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
04641655190
Manufacturer
Roche Diagnostics GmbH
Sandhofer Strasse 116
Mannheim
68305
GERMANY
B. Elecsys Total PSAAll lots
04481798190
Manufacturer
Roche Diagnostics GmbH
Sandhofer Strasse 116
Mannheim
68305
GERMANY
A. Elecsys AFPAll lots

Introduction

Recalled Products

A. Elecsys AFP
B. Elecsys Total PSA

Reason

1) An error was detected in the Cobas E602 analyzer application file version 103 for both kits sizes of Elecsys AFP assay. This file version was inadvertently encoded with a 4 points determination for each calibrator level, when the correct determination is 2-fold.
2) There is a reduced onboard stability for the 100 tests kit size with the AFP application file version 104 and the TPSA application file version 103. The onboard stability of the 100 tests kit was harmonized with the 200 tests kit to the stability claim of (28 days) to allow the use of both reagent kit sizes in parallel on the Cobas E602 with the new software version 04-01.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime