1) An error was detected in the Cobas E602 analyzer application file version 103 for both kits sizes of Elecsys AFP assay. This file version was inadvertently encoded with a 4 points determination for each calibrator level, when the correct determination is 2-fold.
2) There is a reduced onboard stability for the 100 tests kit size with the AFP application file version 104 and the TPSA application file version 103. The onboard stability of the 100 tests kit was harmonized with the 200 tests kit to the stability claim of (28 days) to allow the use of both reagent kit sizes in parallel on the Cobas E602 with the new software version 04-01.