Health Products

Allset Gold DPB1 High Resolution SSP and DPB1 SSP Unitray 12 Tests, DPB1 SSP Unitray with Taq 12 Test (2013-12-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
451606D
Manufacturer
Life Technologies Corporation (Invitrogen)
9099 North Deerbrook Trail
Brown Deer
53223
Wisconsin
UNITED STATES
B. DPB1 SSP Unitray 12 Tests, DPB1 SSP Unitray with Taq 12 TestMore than 10 numbers, contact manufacturer 
54070D
Manufacturer
Life Technologies Corporation (Invitrogen)
9099 North Deerbrook Trail
Brown Deer
53223
Wisconsin
UNITED STATES
A. Allset Gold DPB1 High Resolution SSP008 1073448

Introduction

Recalled Products

A. Allset Gold DPB1 High Resolution SSP
B. DPB1 SSP Unitray 12 Tests, DPB1 SSP Unitray with Taq 12 Test

Reason

There are three reactivity updates with respect to these kits:
1. In primer mix DPB004 (lane 4), the reactivity for the DPB1*31:01 has been updated from positive to negative. Prior to this update, in the presence of the DPB1*31:01 allele, users may have obtained a negative reaction leading to a no-type result.
2. In primer mix DPB027B (lane 27) and DPB039A (lane 39), the reactivity for the DPB1*88:01 allele has been updated from positive to negative. Prior to this update, in the presence of the DPB1*88:01 allele, users may have obtained a negative reaction leading to a no-type result.
3. In primer mix DPB039A (lane 39), the reactivity for the DPB1*29:01 allele has been updated from positive to negative. Prior to this update, in the presence of the DPB1*29:01 allele, users may have obtained a negative reaction leading to a no-type result.