Health Products

Samaritan Pad SAM 500P (2014-02-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Pad 500P
Manufacturer
HeartSine Technologies, Ltd.
Canberra House, 203 Airport Road West,
Belfast, Northern Ireland
BT3 9ED
UNITED KINGDOM
Samaritan Pad SAM 500P 10B0010001 to 14B00461703

Introduction

Recalled Products

Samaritan Pad SAM 500P 

Reason

The software in the Samaritan Pad 500P may miscalculate the CPR rate of compression per minute being administered to the patient. The rescuer may, therefore, be incorrectly advised by the device to "push slower" when, in fact, the compression per minute rate is at an acceptable level.