Health Products

Vitek MS-CHCA Matrix and Vitek MS-DS Target Slides (2014-02-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Not applicableB. Vitek MS-DS Target Slides
Manufacturer
Biomerieux SA
chemin de l'Orme
Marcy l'Etoile
69280
FRANCE
410893
Not applicableA. Vitek MS-CHCA Matrix
Manufacturer
Biomerieux SA
chemin de l'Orme
Marcy l'Etoile
69280
FRANCE
411071

Introduction

Recalled products
  1. Vitek MS-CHCA Matrix
  2. Vitek MS-DS Target Slides

Reason

It has been confirmed that one of the positive QC strains, Enterococcus faecalis ATCC 19433, was not reproducible when tested on Vitek MS following culture on the Trypcase Soy Agar + 5% sheep blood (TSS) media. During development, this Enterococcus faecalis ATCC 19433 strain was tested after culture on the Colombia Agar + 5% sheep blood media. Therefore, the culture of this strain on the Trypcase Soy Agar + 5% sheep blood (TSS) media was never validated.