Health Products

Axiom Vertix MD Trauma (2014-03-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Serial numbers:A. Axiom Vertix MD Trauma
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
8908290

Introduction

Recalled Products
  1. Axiom Vertix MD Trauma

Reason

In rare cases, steel cables inside the lifting column of the system can fail without triggering the safety lock. This can result in the system U-arm dropping down unexpectedly during movement in the vertical direction. Since the U-arm weight is approximately 100 kg, failure of the steel cables may result in serious injury to patients.