Health Products

Kodak 2100 and 2200 Intraoral X-Ray System (2014-03-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturerB. Kodak 2200 Intraoral X-Ray System
Manufacturer
Carestream Health
150 Verona Street
Rochester
14608
New York
UNITED STATES
5153598
More than 10 numbers, contact manufacturerA. Kodak 2100 Intraoral X-Ray System
Manufacturer
Carestream Health
150 Verona Street
Rochester
14608
New York
UNITED STATES
5153622

Introduction

Recalled Products

A. Kodak 2100 Intraoral X-Ray System
B. Kodak 2200 Intraoral X-Ray System

Reason

A potential problem which could result in an early failure of the scissor arm to bracket connection. The failure can be identified by observing the physical condition of the scissor arm bracket. The plastic covers around the bracket typically show signs of fatigue before the bracket actually breaks. This can be an early indicator of a potential problem.