A. Advia Centaur CP System - Procaclitonin (PCT) Assay
10378884
Introduction
Recalled Products
A. Advia Centaur CP System - Procaclitonin (PCT) Assay
B. Advia Centaur XP System -Procalcitonin (PCT) Assay
C. Advia Centaur System - Procalcitonin (PCT) Assay
Reason
Siemens Healthcare Diagnostics received customer feedback regarding the clinical utility of Advia Centaur Brahms Procalcitonin (PCT) quality control level 1. Quality control level 1 currently has a range of plus or minus 50% of the target. Siemens internal investigation confirmed that this range was incorrectly set. Siemens has revised the range of Advia Centaur Brahms Procalcitonin (PCT) quality control level 1 to plus or minus 30% of the target. A review of test data confirmed the assay is capable of achieving a narrower range of plus or minus 30% of the target.