Intellispace Portals - DX and HX (2014-03-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
850001
Manufacturer
Philips Medical Systems (Cleveland) Inc.
3860 NORTH FIRST STREET,
SAN JOSE, CALIFORNIA
95134
UNITED STATES
B. Intellispace Portal HX4598 001 30491
87004
Manufacturer
Philips Medical Systems (Cleveland) Inc.
3860 NORTH FIRST STREET,
SAN JOSE, CALIFORNIA
95134
UNITED STATES
A. Intellispace Portal DX4598 001 31131

Introduction

Recalled Products

A. Intellispace Portal DX
B. Intellispace Portal HX

Reason

During Spect reconstructions using attenuation correction and scatter correction, no scatter correction is being applied in the Autospect Pro application on Intellispace portal with software version 5 and 6. Resolution recovery is not applied correctly in Spect reconstructions using Astonish in Autospect Pro on Intellispace portal unless the data is acquired with VXGP collimators.