A. Intellispace Portal DX
B. Intellispace Portal HX
Reason
During Spect reconstructions using attenuation correction and scatter correction, no scatter correction is being applied in the Autospect Pro application on Intellispace portal with software version 5 and 6. Resolution recovery is not applied correctly in Spect reconstructions using Astonish in Autospect Pro on Intellispace portal unless the data is acquired with VXGP collimators.