Endo-Scrub2 - Sheath and Tubing Set (2014-03-13) Mar 13, 2014 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Brand name Companies Model or catalog number More than 10 numbers, contact manufacturer B. Endo-Scrub 2 Tubing Set Manufacturer Medtronic Xomed Inc. 6743 Southpoint Drive North Jacksonville 32216 Florida UNITED STATES 1912030 More than 10 numbers, contact manufacturer A. Endo-Scrub 2 Sheath Manufacturer Medtronic-Xomed Inc. 6743 Southpoint Drive North Jacksonville 32216 Florida UNITED STATES More than 10 numbers, contact manufacturer
Recalled Products
A. Endo-Scrub 2 Sheath
B. Endo-Scrub 2 Tubing Set
The luer connector on the end of the tubing that attaches to the sheath may break off, either prior to or during use. Once the connector breaks, the flow of saline to the sheath either ceases or is reduced and leakage occurs at the fracture site.