EnGen Laboratory Automation System (2014-03-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
JIM340432EnGen Laboratory Automation System
Manufacturer
Thermo Fisher Scientific Oy
P.O. Box 100, Ratastie 2
Vantaa
01621
FINLAND
EnGen

Introduction

Recalled Products

EnGen Laboratory Automation System

Reason

Ortho Clinical Diagnostics internal testing has determined that when a sample requires follow-up action, such as dilution or retesting due to a mechanical issue, it is possible for that sample to be routed to the sample storage drawer prior to completion of the necessary pending action. A delay in patient test result(s) reporting could occur. This affects engen laboratory automation systems configured with Data Innovations, LLC Instrument Manager V8.06.