C. Immulite 1000 System - Thyroid Autoantibody Assay
LKTG1
244
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
A. Immulite System - Anti-TG AB
LKTG1
Introduction
Recalled Products
Immulite System - Anti-TG AB
Immulite 2000 System - Anti-TG AB
Immulite 1000 System - Thyroid Autoantibody Assay
Reason
Siemens Healthcare Diagnostics has confirmed that the Intra-Assay and Inter-Assay Precision of the Immulite/Immulite 1000 and Immulite/Immulite 2000 XPI Anti-TG AB assays does not align with the representative data published in the instructions for use (IFU). Siemens has conducted a "verification of precision performance" study to assess precision. While the concentrations did not exactly match those published in the IFU, the study provides verification that the current precision performance is not in line with the IFU.