Health Products

Immulite and Immulite 2000 Anti-Tg AB Systems and Immulite 1000 System - Thyroid Autoantibody Assay (2014-03-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
560
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
B. Immulite 2000 System - Anti-TG ABL2KTG2
244
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
C. Immulite 1000 System - Thyroid Autoantibody AssayLKTG1
244
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
A. Immulite System - Anti-TG ABLKTG1

Introduction

Recalled Products
  1. Immulite System - Anti-TG AB
  2. Immulite 2000 System - Anti-TG AB
  3. Immulite 1000 System - Thyroid Autoantibody Assay

Reason

Siemens Healthcare Diagnostics has confirmed that the Intra-Assay and Inter-Assay Precision of the Immulite/Immulite 1000 and Immulite/Immulite 2000 XPI Anti-TG AB assays does not align with the representative data published in the instructions for use (IFU). Siemens has conducted a "verification of precision performance" study to assess precision. While the concentrations did not exactly match those published in the IFU, the study provides verification that the current precision performance is not in line with the IFU.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime