Discovery MR750 and MR450 MR Systems and Optima MR450W MRI Systems (2014-03-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
D. Optima MR450W MRI System - Main Unit - 32CHM7000WC
Not applicable
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
E. Optima MR450W MRI System - Main Unit - GEMM7000ZJ
Not applicable
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
B. Discovery MR450 1.5 T MR System - MagnetM7000BB
Not applicable
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
C. Optima MR450W MRI System - Main Unit - 16CHM7000WA
Not applicable
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
A. Discovery MR750 MR System - Main UnitM7000BH

Introduction

Recalled Products
  1. Discovery MR750 MR System - Main Unit
  2. Discovery MR450 1.5 T MR System - Magnet
  3. Optima MR450W MRI System - Main Unit - 16CH
  4. Optima MR450W MRI System - Main Unit - 32CH
  5. Optima MR450W MRI System - Main Unit - GEM

Reason

The gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance may lead to increasing heat during scanning, which, in turn, can cause material in the rear endbell enclosure to overheat, generate smoke in the scan room and cause some localized melting in the endbell.