Discovery MR750W 3.0T MRI System (2014-03-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
GE Healthcare Japan Corporation
7-127, Asahigaoka, 4-Chome,
Hino-shi, Tokyo
191-8503
JAPAN
Discovery MR750W 3.0T MRI SystemM7000NA

Introduction

Recalled Products

Discovery MR750W 3.0T MRI System

Reason

Gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance can lead to increasing heat during scanning which, in turn, can cause material in the rear endbell enclosure to overheat, generate smoke in the scan room, and cause some localized melting of the endbell.