As a result of investigation into complaints related to the treatment table (TXT) that is part of the linear accelerators, the following safety issues were identified:
In very rare instances, the actual table position may significantly deviate from the position at the table side control and the linac control console, but no interlock occurs.
In sporadic cases the following issue for the table axis lateral, longitudinal and isocentric rotation occurs:
After a manual table movement with released brake an immediate automatic movement for the same table axis is not possible.