Health Products

Ingenuity CT System (2014-03-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
320148
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
A. Ingenuity CT System728323

Introduction

Recalled products
  1. Ingenuity CT System

Reason

Philips has become aware of the following two (2) potential issues that may result in the subframe of the table top becoming free floating and causing unintended, horizontal motion. The two potential issues are:
  1. Service latch may not be secured.
  2. Service latch threaded rod may be fractured.