Health Products

Vitros Chemistry DIluent and Vitros Chemistry FS Specialty Diluent Pack 3 (2014-03-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
01-3266
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
B. Vitros Chemistry FS Diluent Pack 36801754
F3168
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
A. Vitros Chemistry Specialty Diluent855 9825

Introduction

Recalled products
  1. Vitros Chemistry Specialty Diluent
  2. Vitros Chemistry FS Diluent Pack 3

Reason

Ortho Clinical Diagnostics (OCD) internal testing has confirmed the potential for lower than expected Vitros C-Reactive Protein (CRP) results from samples diluted with Vitros specialty diluent or Vitros FS diluent pack 3. When using the affected lots, ocd found that results using Vitros CRP slides may be negatively biased by approximately 51% using a 3x dilution factor.