Brilliance CT Systems and Ingenuity CT System (2014-04-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
320123
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
H. Ingenuity CT System728323
6024
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
F. Brilliance CT System (16P)453567005721
200071
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
G. Brilliance ICT SP System - Main Unit728311
100046
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
D. Brilliance ICT System - System728306
4010
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
E. Brilliance 6/10/16 Slice CT System4535 670 73181
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
B. Brilliance CT Big Bore System - Main Unit4535 670 88051
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
C. Brilliance 64 Slice CT System - Main Unit4550 110 00921
9164
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
A. Brilliance 40/64 Slice CT System4550 110 04011

Introduction

Recalled products
  1. Brilliance 40/64 Slice CT System
  2. Brilliance CT Big Bore System - Main Unit
  3. Brilliance 64 Slice CT System - Main Unit
  4. Brilliance ICT System - System
  5. Brilliance 6/10/16 Slice CT System
  6. Brilliance CT System (16P)
  7. Brilliance ICT SP System - Main Unit
  8. Ingenuity CT System

Reason

Philips has become aware that if the service latch is not properly secured during servicing of the device, the table top's subframe becomes free floating causing unintended, horizontal motion. To date, there have been no reported injuries or deaths resulting from this issue.