Health Products

Unicel DXC 600, DXC 800, and DXC 600I Synchron Access Clinical Analyzer (2014-03-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2342
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B. Unicel DXC 8004764
1581
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
C. Unicel DXC 600I Synchron Access Clinical Analyzer4767
2340
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A. Unicel DXC 6004764

Introduction

Recalled Products
  1. Unicel DXC 600
  2. Unicel DXC 800
  3. Unicel DXC 600I Synchron Access Clinical Analyzer

Reason

Beckman Coulter has confirmed that connector pins in the valves used in the manufacture of the bubble generator assembly cuvette wash manifold and fluid distribution manifold are not crimped per Beckman Coulter workmanship standards.