Health Products

Mobilediagnost WDR (2014-01-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsMobilediagnost WDR
Manufacturer
Philips Medical Systems DMC GmbH
Rontgenstrabe 24
Hamburg
22335
GERMANY
9890 010 8952X

Introduction

Recalled Products

Mobilediagnost WDR

Reason

Philips has become aware that under very specific error conditions the system may execute an unintended exposure:

  • the hand switch is erroneously stuck in pressed position, while the system is switched off,


  • the system is powered up with the hand switch in that erroneous condition,


  • the user logs in, selects a patient for examination, gets a green ready light in the examination context, if the tube arm is then moved out of the zero position, the system is doing an exposure.


Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime