Philips has become aware that under very specific error conditions the system may execute an unintended exposure:
the hand switch is erroneously stuck in pressed position, while the system is switched off,
the system is powered up with the hand switch in that erroneous condition,
the user logs in, selects a patient for examination, gets a green ready light in the examination context, if the tube arm is then moved out of the zero position, the system is doing an exposure.