Intra-Aortic Balloon Pumps - Systems 98, 98XT, CS100, CS100I, and CS300 (2014-03-24) Mar 24, 2014 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Brand name Companies Model or catalog number More than 10 numbers, contact manufacturer. D. CS100 Intra Aortic Balloon Pump Manufacturer Datascope Corp. 1300 Macarthur Blvd. Mahwah 07430 New Jersey UNITED STATES 0998-00-3013-XX More than 10 numbers, contact manufacturer. E. CS100I Intra Aortic Balloon Pump Manufacturer Datascope Corp. 1300 Macarthur Blvd. Mahwah 07430 New Jersey UNITED STATES 0998-00-0446H53 More than 10 numbers, contact manufacturer. B. CS300 Intra Aortic Balloon Pump Manufacturer Datascope Corp. 1300 Macarthur Blvd. Mahwah 07430 New Jersey UNITED STATES 0998-00-3023-XX More than 10 numbers, contact manufacturer. C. System 98XT Intra Aortic Balloon Pump Manufacturer Datascope Corp. 1300 Macarthur Blvd. Mahwah 07430 New Jersey UNITED STATES 0998-00-0479-XX More than 10 numbers, contact manufacturer. A. System 98 Intra-Aortic Balloon Pump Manufacturer Datascope Corp. 1300 Macarthur Blvd. Mahwah 07430 New Jersey UNITED STATES 0998-00-0446-XX
Recalled products
System 98 Intra-Aortic Balloon Pump
CS300 Intra Aortic Balloon Pump
System 98XT Intra Aortic Balloon Pump
CS100 Intra Aortic Balloon Pump
CS100I Intra Aortic Balloon Pump
The fan assembly for specific system 98/98XT, CS100/CS100I and CS300 IABPs could potentially contain a misshapen retaining ring. This retaining ring could disengage within the fan assembly, causing the fan to stop rotating, which causes the power supply to overheat and the IABP to shut down without any visual or audible warning.