Health Products

Femoral Arterial Cannula - Fem-Flex II and Duraflo II Fem-Flex II (2014-04-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Edwards Lifesciences LLC
1 Edwards Way
Irvine
92614
California
UNITED STATES
B. Duraflo II Fem-Flex II Femoral Access CannulaDIIFEMII016A
More than 10 numbers, contact manufacturer
Manufacturer
Edwards Lifesciences LLC
1 Edwards Way
Irvine
92614
California
UNITED STATES
A. Fem-Flex II Femoral Arterial CannulaFEMII016A

Introduction

Recalled Products

A. Fem-Flex II Femoral Arterial Cannula
B. Duraflo II Fem-Flex II Femoral Access Cannula

Reason

Edwards Lifesciences has received five complaints in the last four months involving the separation of dilator tips on specific models of Fem-Flex II Arterial Cannula sizes 16, 18, and 20 French. Some of the reported events involved separation of the tip of the dilator and required retrieval through an interventional radiologic procedure.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime