Health Products

Compact Airway Module (2014-04-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
GE Healthcare Finland OY
Kuortaneenkatu 2,
Helsinki
00510
FINLAND
A. Compact Airway ModuleE-CAIO-00-EN

Introduction

Recalled products
  1. Compact Airway Module

Reason

Loosening of the gas sampling pump hex screw in the module pump unit may lead to pump failure that may cause loss of respiratory airwway gas parameter monitoring. If this issue occurs, the monitor will show an alarm note on monitor screen of either "low gas sample flow" or "check sample gas out".
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime