Health Products

Hexapod Evo-RT-Couchtop (2014-04-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Medical Intelligence Medizintechnik Gmbh
Robert-Bosch-Strasse 8
Schwabmunchen
86830
GERMANY
A. Hexapod Evo-RT-CouchtopP10603-700

Introduction

Recalled products
  1. Hexapod Evo-RT-Couchtop

Reason

The Hexapod EVO module is fixed to the base table with six screws. If the screws are not properly tightened during installation, the screws may become loose over time. If the screws become loose, there is a potential that the patient could fall off the couchtop.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime