Health Products

Hamilton - G5 Ventilator (2014-04-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All with software versions 2.00 to  2.31A. Hamilton - G5 Ventilator
Manufacturer
Hamilton Medical AG
Via Crush 8
Bonaduz
7402
SWITZERLAND
159001

Introduction

Recalled products
  1. Hamilton - G5 Ventilator

Reason

Investigations shown that ventilation and alarms of a Hamilton-G5 ventilator can be supressed unintentionally after the activation of a suctioning maneuver. Can occur in any patient group (Neonatal, Pediatric and Adult)