Giraffe and Panda Warmers (2014-04-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Ohmeda Medical, A Division of Datex Ohmeda Inc.
8880 Gorman Road
Laurel
20723
Maryland
UNITED STATES
B. Panda Warmer2063568-001
More than 10 numbers, contact manufacturer
Manufacturer
Ohmeda Medical, A Division of Datex Ohmeda Inc.
8880 Gorman Road
Laurel
20723
Maryland
UNITED STATES
A. Giraffe WarmerM1118179

Introduction

Recalled Product

A. Giraffe Warmer

B. Panda Warmer

Reason

There may be a loss of internal communication that manages the user-adjusted Nellcor SPO2 alarm settings of the affected Giraffe and Panda Ires warmer systems with Nellcor SPO2. The alarms may not activate as expected, which can result in false positive or false negative saturation and pulse rate alarm notifications to the caregiver. The patient's oxygen saturation and pulse rate values are accurate as displayed however, the displayed user-set alarm limits may be different than the actual alarm limits used for alarm activation. All other clinical functionality of the warmer is unaffected.