Health Products

Maquet Critical Care Servos U and N (2014-04-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
1007B. Servo-N
Manufacturer
Maquet Critical Care AB
Rontgenvagen 2,
Solna
17154
SWEDEN
6688600
20061A. Servo-U
Manufacturer
Maquet Critical Care AB
Rontgenvagen 2,
Solna
17154
SWEDEN
6694800

Introduction

Recalled Products

A. Servo-U

B. Servo-N

Reason

We have had one occurrence of a disconnection from the ventilator tubing, which was not detected by the ventilator, (i.e. No alarms were triggered.) This absence of alarms may occur with system version 1.00.05 in "invasive ventilation modes", in "adult and pediatric patient categories" and with the low peep alarm turned off by setting it to zero. This absence of alarms does not occur in the neonatal patient category, or in non-invasive modes. This absence of alarms does not occur in system version 1.00.06, which is readily available.