H. Advia Centaur XP System - Vitamin D Total (VITD)
10491994
15945020
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
F. Advia Centaur System - Procalcitonin (PCT) Assay
10378883
More than 10 numbers, contact manufacturer
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
G. Advia Centaur System-Vitamin D Total (VITD) Reagent
10491994
15945020
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
D. Advia Centaur CP System - Procalcitonin (PCT) Assay
10378883
15945020
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
E. Advia Centaur XP System - Procalcitonin (PCT) Assay
10378883
More than 10 numbers, contact manufacturer
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
B. Advia Centaur CP System - TSH3 - Ultra (TSH3-Ul) Assay
06491080
More than 10 numbers, contact manufacturer
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
C. Advia Centaur XP - TSH3 - Ultra (TSH3-Ul) Assay
06491080
More than 10 numbers, contact manufacturer
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
A. Advia Centaur System - TSH3 - Ultra (TSH3-Ul) Assay
06491080
Introduction
Recalled Products
A. Advia Centaur System - TSH3 - Ultra (TSH3-Ul) Assay
B. Advia Centaur CP System - TSH3 - Ultra (TSH3-Ul) Assay
C. Advia Centaur XP - TSH3 - Ultra (TSH3-Ul) Assay
D. Advia Centaur CP System - Procalcitonin (PCT) Assay
E. Advia Centaur XP System - Procalcitonin (PCT) Assay
F. Advia Centaur System - Procalcitonin (PCT) Assay
G. Advia Centaur System-Vitamin D Total (VITD) Reagent
H. Advia Centaur XP System - Vitamin D Total (VITD)
Reason
Siemens Healthcare Diagnostics has confirmed that samples containing fluorescein may show interference with the Advia Centaur systems TSH3 Ultra, Vitamin D and the Advia Centaur Brahms Procalcitonin assays. Samples containing fluorescein can produce falsely depressed values when tested with Advia Centaur TSH3 ultra assay. Samples containing fluorescein can produce falsely depressed values when tested with the Advia Centaur Brahms Procalcitonin assay. Samples containing fluorescein can produce falsely elevated values when tested with the Advia Centaur Vitamin D assay.