Health Products

Apex Daily Wear Contact Lenses (2014-04-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Material batch number:  4291 
Manufacturer
Cardinal Contact Lens Inc.
38 Cowansview Rd., Unit 6 and 7
Cambridge
N1R 7N3
Ontario
CANADA
Apex Daily Wear Contact LensesAspherix Cone

Introduction

Recalled Products

Apex Daily Wear Contact Lenses

Reason

A number of units produced from ONSI-56 batch number 4291 have demonstrated dimensional instability in the form of base curve warpage and/or steepening occurring in post production, potentially resulting in decreased visual performance.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime