Advanced Perfusion System 1 - Base Unit (2013-04-23) Apr 23, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number 0011-1137 Manufacturer Terumo Cardiovascular Systems Corp. 6200 Jackson Road Ann Arbor 48103 Michigan UNITED STATES Advanced Perfusion System 1 - Base Unit 801763
Recalled Products
Advanced Perfusion System 1 - Base Unit
System malfunction caused by an inadequate crimp quality on internal cables creating loss or fluctuation of internal power on one side of the system's base.