Health Products

Advanced Perfusion System 1 - Base Unit (2013-04-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
0011-1137
Manufacturer
Terumo Cardiovascular Systems Corp.
6200 Jackson Road
Ann Arbor
48103
Michigan
UNITED STATES
Advanced Perfusion System 1 - Base Unit 801763

Introduction

Recalled Products

Advanced Perfusion System 1 - Base Unit 

Reason

System malfunction caused by an inadequate crimp quality on internal cables creating loss or fluctuation of internal power on one side of the system's base.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime