Architect System - Hemoglobin A1C Calibrators(2014-05-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
45063UQ12
Manufacturer
Abbott Laboratories Diagnostcis Division
100 Abbott Park Road,
Abbott Park
60064
Illinois
UNITED STATES
Architect System - Hemoglobin A1C Calibrators4P52-02

Introduction

Recalled Products

Architect System - Hemoglobin A1C Calibrators

Reason

HBA1C calibrators LN 4P52-02, lot 45063UQ12 may contain a value sheet for a different HBA1C calibrator lot. If calibrator lot 45063UQ12 is used in the calibration of the HBA1C assay using values from an incorrect calibrator value sheet then all results, including quality control and patient, may be falsely elevated.