Mammomat Inspiration System - Acquisition Workstation (2014-05-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
1146A. Mammomat Inspiration System - Acquisition Workstation
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
101 43 539

Introduction

Recalled products
  1. Mammomat Inspiration System - Acquisition Workstation

Reason

In rare cases, the holder of the PC monitor can break and this will cause an unstable monitor. Since the monitor weight is approximately 8kg, it is assumed that the monitor drops down on the table or on the floor, this can result in a serious injury of the hand or the foot or the operator. Only a specific range of systems are affected by this recall. The serial number located in the letter is for a specific system component, not the mammoth inspiration system in which it is located.