Health Products

Trigen Titanium Nail System - Intertan Short Nails (2014-04-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
13LT36306
Manufacturer
Smith and Nephew, Inc.
1450 Brooks Road
Memphis
38116
Tennessee
UNITED STATES
Trigen Titanium Nail System - Intertan Short Nails71675201

Introduction

Recalled products

Trigen Titanium Nail System - Intertan Short Nails

Reason

Smith & Nephew has initiated a product recall of several batches of the Trigen Intertan nail due to a manufacturing error, which resulted in the set screw being set too deep in the affected nails. During implantation, when the proximal femur is being reamed to accept the lag and compression screw, the reamer may impinge the set screw, causing potential breakage of the reamer.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime