Recalled Products
MacroLyte Premie Dispersive Electrode
Through the Conmed corporation complaint handling system, Conmed received six (6) complaints regarding sparking, no output and burning at the cord set connection. The initial complaint was received on 04/11/2014. The complaints were for the gel possibly making contact with an internal staple and this may cause no output, sparking and heat buildup between the staple and foil contact surface. One patient in Japan was reportedly to have received a burn. Conmed confirmed instances where the gel contact with the internal staple causes the staple to corrode over time. Corroded staples may fail upon device activation, causing a burn through the foam which covers the staple.